The Lepu Medical Grade PC-303 Telehealth All-in-One Vital Signs Monitor has been approved by FDA and CE. It can measure Non-Invasive Blood Pressure (NIBP), Blood Glucose/Glucose, SpO2 (Oxygen Saturation), ECG (Electrocardiogram), PR (Pulse Rate), and Temperature via its 4.3" LCD display. Suitable for home, doctor, clinic, nursing staff, medical institution, medical facility environment, and outdoor measurement.
● PC-303 is equipped with professional ECG monitoring equipment to provide adequate medical services to patients at home through telemedicine.
4.3-inch display of SpO2, Pulse Rate, NIBP(Non-invasive Blood Pressure), Temperature , ECG, and Blood glucose.
Real-time display for both IOS and Andriod systems via App.
PC data management software for data export and analysis.
Optional Briefcase with Andriod tablet for outdoor measurement.
Communication protocol and SDK for Android, IOS, and Windows for further development.
Equipped with 1/7 leads ECG for optional, real-time waveform display to meet a variety of doctor requirements.
Specifications | |
NIBP | |
Measuring method | Oscillometric Technique |
Measuring range | 0mmHg~300mmHg |
Maximal standard deviation | ≦ 8mmHg |
Inflation time | <20s (typical adult cuff) |
Overpressure protection limit | 300mmHg |
Measuring range | SYS: 30mmHg~270mmHg |
DIA: 20mmHg~235mmHg | |
MAP: 10mmHg~220mmHg | |
Temperature | |
Measuring range | 32 ℃ ~43 ℃ |
Measuring accuracy | ±0.2 ℃ |
Response time | ≦ 5s |
Probe | Infrared ear probe |
SpO2 | |
Technique | Dual-wavelength LED |
SpO2 measuring range | 0%~100% |
SpO2 measuring accuracy | ±3% (range from 70%~100%) |
PR measuring range | 30bpm~250bpm |
PR measuring accuracy | ±2bpm |
Sensor type | Adult, Pediatric |
Blood Glucose | |
Measuring range | 20 - 600 mg/100ml (1.1-33.3mmol /L) |
Measuring time | 5s |
Sample volume | ≥ 0.7 microlitre |
Measuring method | Electrochemical biosensor method |
ECG | |
ECG electrodes | 3-lead, 5-lead |
Input impedance | ≥50MΩ, 10Hz |
Input signal range | 10mV (peak-to-valley value) |
Common-mode Rejection Ratio | ≥120dB |
Bandwidth | 0.05 ~ 40 Hz |
Heart Rate | |
Measurement Range | 30 ~ 250 bpm |
Measurement Error | ±2bpm, ±2%, whichever is larger |
Resolution | 1 bpm |